您當(dāng)前的位置: 首頁-產(chǎn)品展示-HR2000+ES 光譜儀
HR2000+ES 光譜儀
2019/08/07
HR2000+ES High-resolution Spectrometer
for Enhanced Sensitivity
The HR2000+ES Spectrometer is a preconfigured system that integrates a high-resolution optical bench, a powerful 2-MHz analog-to-digital (A/D) converter, programmable electronics, a 2048-element CCD-array detector, and a high-speed USB 2.0 port. This innovative combination produces and extremely fast spectrometer and provides resolution to ~1.33 nm (FWHM).
The HR2000+ES has a slightly larger entrance slit (10 microns) and an L2 quartz collection lens to allow for extended measurement applications including:
Reflection (emissive and reflective color)
Absorbance
Irradiance
Transmission
The HR2000+ES allows users to capture and store a full spectrum into memory every millisecond (that's 1,000 full spectra every second) when the spectrometer is interfaced to a computer via a USB 2.0 port. The HR2000+ES is perfect for chemical, biochemical and other applications where fast reactions need to be monitored and fine spectral features need to be resolved.
1,000 Full Spectra/Second
Programmable Microcontroller
High-resolution Applications
HR2000+ES Sample Spectrum
High-resolution "HR" Optical Bench
Plug and Play Operation
Configuration
Pricing
Specifications
1,000 Full Spectra/Second
The HR2000+ES utilizes an onboard, 2-MHz A/D converter, which allows you to capture and transfer one full spectrum into memory every millisecond when the spectrometer is interfaced to a PC via the USB port.
Programmable Microcontroller
The HR2000+ES has an onboard programmable microcontroller that provides flexibility in controlling the spectrometer and accessories. Through a new 30-pin connector, you can implement all operating parameters in the software, such as controlling external light sources, creating processes and routines and retrieving data from external devices. The HR2000+ES gives you access to 10 user-programmable digital I/Os for interfacing to other equipment; one analog input and one analog output; and a pulse generator for triggering other devices. (Programming the I/Os requires SpectraSuite Spectroscopy Operating Software.)
High-resolution Applications
The HR2000+ES is ideal for applications where fast reactions need to be monitored and high resolution is necessary, such as protein dynamics.
HR2000+ES Sample Spectrum
This power-up data for our LS-1 Tungsten Halogen Light Source was taken by an HR2000+ at 2-millisecond intervals. The graph shows the tremendous amount of data generated with the HR2000+s acquisition rate speed of 1000 spectra per second.
High-resolution "HR" Optical Bench
The exceptional ~1.33 nm optical resolution of the HR2000+ES is achieved by using the Ocean Optics High-resolution Optical Bench ?a 25.4-mm diameter, 101.6-mm focal length, symmetrical crossed Czerny-Turner optical design.
Plug and Play Operation
The HR2000+ES interfaces to a computer via USB 2.0 or RS-232 serial port. When using the serial port, the HR2000+ES requires a single 5-volt power supply (not included). Data unique to each spectrometer are programmed into a memory chip on the HR2000+; SpectraSuite Spectrometer Operating Software reads these values for easy setup and hot swapping among computers
Configuration
Equipment | Details |
Spectrometer | HR2000+ |
Grating | HC1 |
Detector | DET2B-200-1100 |
Slit | SLIT-10 |
Collection Lens | L2 |
Pricing
Item | Description | Price |
HR2000+ES | HR2000+ES User-configured Spectrometer. system. Includes SpectraSuite and an ASP. | $6044 |
Specifications
Physical | |
Dimensions: | 148.6 mm x 104.8 mm x 45.1 mm |
Weight: | 570 grams |
Detector | |
Detector: | Sony ILX511B linear silicon CCD-array |
Detector range: | 200-1100 nm |
Pixels: | 2048 pixels |
Pixel size: | 14 μm x 200 μm |
Pixel well depth: | ~62,500 electrons |
Sensitivity: | 75 photons/count at 400 nm; 41 photons/count at 600 nm |
Optical Bench | |
Design: | f/4, Symmetrical crossed Czerny-Turner |
Focal length: | 101.6 mm input and output |
Entrance aperture: | 10 m wide slit |
Grating: | HC-1 provides 200-1100 nm range |
Detector collection lens: | L2 |
OFLV filter: | OFLV 200-1100 nm |
UV enhanced window: | UV2 |
Fiber optic connector: | SMA 905 to 0.22 numerical aperture single-strand optical fiber |
Spectroscopic | |
Wavelength range: | 200-1100 nm (Grating dependent) |
Optical resolution: | ~1.33-6.8 nm FWHM |
Signal-to-noise ratio: | 250:1 (at full signal) |
A/D resolution: | 14 bit |
Dark noise: | 12 RMS counts |
Dynamic range: | 2 x 10^8 (system); 1300:1 for a single acquisition |
Integration time: | 1 ms to 65 s (20 s typical max.) |
Stray light: | <0.05% at 600 nm; <0.10% at 435 nm |
Corrected linearity: | >99.8% |
Electronics | |
Power consumption: | 220 mA @ 5 VDC |
Data transfer speed: | Full scans to memory every 1 ms with USB 2.0 port, 15 ms with USB 1.1 port |
Inputs/Outputs: | Yes, 10 onboard digital user-programmable GPIOs |
Analog channels: | One 13-bit analog input; one 9-bit analog output |
Auto nulling: | No |
Breakout box compatibility: | Yes, HR4-BREAKOUT |
Trigger modes: | 4 modes |
Strobe functions: | Yes |
Gated delay feature: | No |
Connector: | 30-pin connector |
Computer | |
Operating systems: | Windows 98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS with serial port |
Computer interfaces: | USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 115.2 K baud |
Peripheral interfaces: | SPI (3-wire); I2C inter-integrated circuit |